Krohne UFM 600T EN Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Équipement Krohne UFM 600T EN. KROHNE UFM 600T EN User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 93
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
Installation and
operating instructions
UFM 600 T
ALTOSONIC
©
KROHNE 01/94 7.30718.32.00
GR
Variable area flowmeters
Vortex flowmeters
Flow controllers
Electromagnetic flowmeters
Ultrasonic flowmeters
Mass flowmeters
Level measuring instruments
Communications engineering
Engineering systems & solutions
Switches, counters, displays and recorders
Heat metering
Pressure and temperature
Vue de la page 0
1 2 3 4 5 6 ... 92 93

Résumé du contenu

Page 1 - ALTOSONIC

Installation andoperating instructionsUFM 600 TALTOSONIC©KROHNE 01/94 7.30718.32.00GRVariable area flowmetersVortex flowmetersFlow controllersElectro

Page 3

B.2 Measuring SystemA microprocessor controls the transducers via an analog/digital interface and calculates the actual flow.The control program is st

Page 4

Fig. B.2 : System diagram11SIGNALPROCESSORMICROPROCESSOR(µP2)KEYBOARDEEPROMRAMPOWERSUPPLYDISPLAYFREQUENCY*OUTPUTCURRENT OUTPUT*RS 232 OUTPUTUP DOWNFLO

Page 5

B.3 EquipmentThe UFM 600 T flowmeter is fitted in a sturdy case.Figure B.3 shows the UFM 600 T with all itemsdisplayed: 1. UFC 600 T signal converter

Page 6

Fig. B.3 : UFM 600 T with all equipment displayed13➉

Page 7

B.4 Controls and ConnectionsFigure B.4 shows the controls and connections of the flowmeter unit.1 Liquid crystal display2 Keyboard (3 keys)3 Power sup

Page 8

Fig. B.4 : UFC 600 T converter159

Page 9

B.5 Transducers and Mountings (see Fig. B.5)Fig. B.5 : Transducers, signal converter and auxillary itemsTwo mounting rails are clamped against the pip

Page 10

B.6 Display and Data StorageDisplay:The following display options are available:- Actual flow rate and direction.- Total positive and negative flow vo

Page 11

B.7 Direct OutputBoth current and frequency output signals are available forcontrol purposes. These can be either analog signals or justflow direction

Page 12 - Fig. B.2 : System diagram

All rights reserved. No part of this publication may be copied or publishedby means of printing, photocopying, microfilm or otherwise without priorwri

Page 14

C.1 Initial Set-UpThe main steps of the installation procedure of the flowmeterare:1 Install the mounting rails on the pipe.2 Program the applicatio

Page 15

C.1.1 Installation of mounting railsMounting Considerations:The pipe section on which the transducers will be clampedmust always be completely filled

Page 16

Mounting Procedure:a Severe rust or thick coatings have to be removed from the pipe where the transducers will be installed.b Slide the securing strap

Page 17

d Use the supplied tape measure to ensure that the rails are fitted diametrically opposite each other (see fig. C.3-1a. ).Distance A1 must equal dista

Page 18

C.1.2 Programming the application parametersAfter the flowmeter has been switched on it will start in themeasuring mode.Pressing twice the key e

Page 19

Input of valuesInput of values outside the range, indicated at the functiondescriptions, will result in a flashing error message; the mini-mum or maxi

Page 20

C.1.3 Installation of the transducersSince the required transducer distance "S" is known now thetransducers can be positioned in the rails:1

Page 21

27NOTE: - The cable connections at the transducers must always point away from each other (fig. C.5).- Make sure that the transducers are not placed o

Page 22

C.1.4 Zero set / Zero point calibrationZero SetAfter leaving the transducer distance function the ZERO SETfunction can be activated.The flowmeter is f

Page 23

ALTOSONICA INTRODUCTION ...6B SYSTEM DESCRIPTION ...

Page 24

Zero Point CalibrationBefore performing the zero point calibration ensure that:- The unit is functioning properly (see low signal marker).- There is n

Page 25

C.2 Start Up MenuC.2.1 Enter start up menuThe Start Up Menu is activated by pressing the key twicefrom the measuring mode or pressing the

Page 26

PIPE MAT.Program the sonic velocity of the pipe material.The following 5 options are preprogrammed (or will reappearafter running 3.5.8 DEFAULT SET, s

Page 27

LINING MAT.(does not appear on display when lining = "NO") Program the sonic velocity of the lining material.The following 5 options are pre

Page 28 - FLOWDIRECTION

according to the user´s requirements (names and sonicvelocities).NOTE: If the sonic velocity is not known it can be estimated by usingthe sonic veloci

Page 29

ZERO SET :Zero point calibration.Select between the factory set default value for the zero pointor perform a zero point calibration for the actual sit

Page 30

C.3 Error DetectionC.3.1 Low signal marker functionThe condition of the input signal from the transducers is indi-cated by a marker in the top left ha

Page 31

C.3.2 Error messagesIf the microprocessor produces an error message a * signappears at the lower left hand side of the display.Enter the QUIT / RESET

Page 32

37STORE: A total number of 350 data blocks arebeen stored. Rest capacity 40 datablocks till data storage is full.FATAL ERROR: Check sum error detected

Page 33

C.4 Data CollectionC.4.1 Data collection procedureSampling Rate / Data Storage CapacityA total number of 390 data blocks can be stored.This number is

Page 34

D PROGRAMMING MODE... 48D.1 General ...

Page 35

Preparation for new measurementsProviding any previous flow data can be discarded, a datacollection session can be started as follows:1 Select the STO

Page 36

C.4.2 Transfer to PC (data and parameters)SoftwareThe floppy disk provided contains programs for the transferand conversion of the measuring data:- GE

Page 37

Data transferFor the transfer of information from the flowmeter to the PCproceed as follows:- Connect the RS 232 cable to the flowmeter and the serial

Page 38

C.5 Quit / Reset MenuThe measuring mode can be exited with the code asdescribed below to enter the Quit/Reset Menu.In this menu two functions can be u

Page 39

RESET:In this submenu three reset options are available:TOTAL + : Reset + totalizer.TOTAL - : Reset - totalizer.RESET DATA : Reset all data blocks.In

Page 40

C.6 Direct Output Ports (Analog (mA) output and frequency / pulse output)C.6.1 General The options available for these direct output ports are descri-

Page 41

Active output short-circuit-proofTerminals 4.1/4.2 : for electromechanical (EMC) orelectronic (EC) totalizersTerminals 4/4.1/4.2: for electronic (EC

Page 42

C.6.5 Connection Diagrams45aALTOSONIC OperationUFM 600 TI:(term.5/6)Factie:EC (term.4/4.1/4.2)FpassiveEC (term.4/4.1)FactiveEMC/EC(term.4.1/4.2)4 4.14

Page 44

C.7 Trouble Shooting1 The flowmeter does not respond after connection tothe main power supply (no display or output messages).- Check fuse.2 The er

Page 45

Fig. B.1 : Measuring principle... 8Fig. B.2 : System diagram...

Page 46

replacements. If 3 or 4 markers remain the selection of a diffe-rent location on the pipe wall is re-commended; some irregularities atthe interior of

Page 47

D.1 GeneralIn the main menu, two Sub Menus can be activated:Sub Menu 2 "TEST":Four function tests can be executed:- Display test- Microproce

Page 48

D.2 How to ProgramPush button functions:move the menu pointer one step to the right toactivate the submenu displayed.either move the menu pointer to t

Page 49

Menu selection:By entering the main menu, the submenu Start Up 1.0.0. willbe on display; use the key to select the submenus TEST2.0.0. or INSTALL

Page 50

If any discrepancies are detected the error messagePARAMERROR is displayed and a separate menu is activat-ed in which the parameters concerned can be

Page 51

D.3 List of Functions2.00 TEST2.1.0 DISPLAY2.1.1 DISP.TEST2.2.0 PROCESSOR2.2.1 8048 TEST2.3.0 CURR. OUTP.2.3.1 TEST I2.4.0 FREQ. OUTP.2.4.1 TEST F3.0.

Page 52

3.4.0 FREQ OUTP. Set up frequency/pulse output.3.4.1 FUNCTION F3.4.2 PULSOUTP.3.4.3 PULSRATE PULS/UNIT3.4.4 T- CONST. F3.5.0 SPEC. FCT. Set special f

Page 53

D.4 Sub Menu: 3.0.0 INSTALLATION 3.1.0 FLOW DATA : set flow parameters3.1.1 PRIM. CORR. : Primary CorrectionThis function allows the user to correct

Page 54

3.1.3 L.F. CUTOFF : Low Flow CutoffTo avoid erroneous measurements at low flow rates the LowFlow Cutoff can be used ( see Fig. D.2).Fig. D.2 : Low flo

Page 55

3.1.4 CUTOFF ON : (only if at 3.1.3 YES wasentered)Program the flow rate level below which all outputs will beswitched off.Enter the value in % of the

Page 57

3.2.0 DISPLAY : set output functionsL.C.displayNOTE: All functions that are programmed to be on display can alsobe saved as programmed in function 3.6

Page 58

3.2.3 UNIT TOTAL : unit for totalizers(only if a totalizer has been set in 3.2.2)Select unit for the totalizers using the key and press to store:

Page 59

3.2.6 CYCL. DISPL. : Cyclic DisplayThe cyclic display option can be used to automaticallychange the function of the display, if more than one variable

Page 60

3.3.0 CURR. OUTP.I : set up analog currentoutput3.3.1 FUNCTION I : Program the analog current outputSelect an output option using the key and pre

Page 61

Fig. D.3 : Bi-directional flow (I) Fig. D.4 : Flow direction indication (I)or signal loss indication (I) Fig.D.5 : Positive flow (I) Fig. D.6 : Nega

Page 62

3.3.2 [I 0 PCT] : Program current output at 0 %flowRange: 00 mA ≤ [I 0 PCT] ≤ 16 mANOTE: This value must be lower than the [I 100 PCT] value !!If th

Page 63

3.4.0 FREQ. OUTP. F : set up frequency / pulse output3.4.1 FUNCTION F : Program the analog fre-quency outputSelect an output option using the

Page 64

Fig.D.9: Positive flow (F)3.4.2 PULSOUTP. : Frequency output modeSelect one of the two frequency output modes using the key and press to stor

Page 65

or signal lost indication (F)3.4.3 PULS/UNIT - PULSRATE(pulse/unit or pulse rate depends on selection in 3.4.2)PULSE/UNIT : Program number of pul-ses

Page 66

3.5.0 SPEC. FCT. : set special functions3.5.1 LANGUAGE : Display languageSelect a language option using the key and pressto store:GB/USA : E

Page 67

System DescriptionThe Altosonic UFM 600 T is an ultrasonic flowmeter which can easily beclamped on existing pipelines with inner diameters between 50

Page 68

3.5.4 FACT. QUANT.If a user defined unit is applied the quantity factor has to beprogrammed.This must be the conversion ratio into m3.In the example w

Page 69

3.5.8 DEFAULT SETIf this function is used all parameters will be reprogrammedto default settings (see Appendix 5).(NOTE: factory settings are not affe

Page 70

3.6.0 COMM. DATA : set RS 232 output /store3.6.1 OUTP. PARAM.With this function all programmed parameters can be sent toa PC via the RS 232 output.NOT

Page 71

3.6.4 INTERVAL (not visible if SEL. COMM is OFF)Program the interval between successive data collections asprogrammed in 3.6.3.Range: 01 - 60 min3.6.5

Page 73

D.5 Sub Menu : 2.0.0 TEST2.1.0 DISPLAY2.1.1 DISPLAY TESTAll individual segments of the liquid crystal display areactivated in a sequence ending in a

Page 74

2.4.0 FREQ. OUTPUT2.4.1 TEST To check the functioning of both the frequency output andthe externally connected instruments and wiring, the frequen-cy

Page 75

D.6 Parameter errorsIf discrepancies in the parameters entered are found themicroprocessor will display the message "PARAMERROR"thus prevent

Page 76

4.4.1 L.F. CUTOFF4.4.2 CUTOFF ON4.4.3 CUTOFF OFF4.5.0 F>10KHZ Output frequency is too high,must be less than 10kHz4.5.1 FULL SCALE4.5.2 PULSOUTP.4.

Page 77

E.1 Mains FuseThe mains fuse is located in the connection compartment.NOTE: ONLY USE FUSES AS INDICATED NEXT TO THE FUSE FITTING TO PREVENT SERIOUS DA

Page 80

F.1 GeneralMeasuring beam 1Diameter range 50 ... ≥ 3000 mm (2” ... ≥120”)Pipe wall thickness ≤ 40 mm ( ≤ 1.6”)Pipe wall materials metal, plastics, c

Page 81

F.2 Transducers & Clamp-on SetTransducers single beam; 2x RS 600Diameters A single clamp-on set suitable for50 mm (2”) to 2000 mm (80”)pipes, inco

Page 82

F.3 Signal ConverterVersion UFC 600 TOutputs fully programmablegalvanically insulated - analog 0 to 20 mA, or 4 to 20 mA, orI[0%] to I[100%] programma

Page 83

Local display: back-lit LCDDisplay functions actual flow rate, forward-, reverse-and sum-totalizers (7 digit), transi-tion time and error messages;eac

Page 84

Power supplyAC power supply: 85 - 264 VACDC power supply: 18 - 32 VDCPower consumption ≤ 10 VA AC≤ 8 W DCHousing Cast aluminium82ALTOSONIC Technical

Page 85

Ambient temp. -25 to +50 °C (-13 to 122°F)Protectioncategory IP 65, acc. to IEC 529 (equivalent to NEMA 4x)Full scale value (Din= inner pipe diameter

Page 86

APPENDIX 1 : Pipe Material Sonic VelocitiesALTOSONIC Appendix84MATERIAL NAME SONIC VELOCITYCarbon Steel c ≤ 0.3% 3064 m/sCarbon Steel c >

Page 87

APPENDIX 2 : Sonic Velocities of Liquids 85LIQUID NAME SONIC VELOCITY at 20°CAcetone 1174 m/sBenzene 1295 m/sCarbon Disulphide 1149 m/sCaster

Page 88

ALTOSONIC Appendix86APPENDIX 3 : Sonic velocity calculation Using Indicated Transit Time StartMeasuringModeCheckindicatedtrans-ducerRestartprocedureIn

Page 89

B.1 Measuring PrincipleFig. B.1 : Measuring principleA sound wave that is sent in the direction of the flow througha medium will travel faster than on

Page 90

APPENDIX 4 : Exponential NotationMost values have to be entered in the exponential notation as will bemade clear by the message displayed.Exponential

Page 91

UFM 600 TAPPENDIX 5 : Default Settings + Example of Output ParametersThe next page shows all parameters after the function Default Set (3.5.8)has been

Page 92

1.0.0. STARTUPDIAMETER1.0000E+2mmPIPE WALL1.0000E+0mmPIPE MAT.3.1700E+3LININGNOLIQUID1.5000E+3WATERFULL SCALE1.0000E+2m3/hrLOCATIONSENS. DIST.4.4329E+

Page 93

UFM 600 TAPPENDIX 6 : Spare PartsKrohne Description Number Type/remarksordering pernumber UFM 600 T230262- 01 Mounting Rail 2 UL 600R (with slots)2312

Commentaires sur ces manuels

Pas de commentaire